Pulse oximetry and capnography are the two most widely used continuous monitoring modalities in acute care nursing. Together they give the nurse a real-time window into two distinct but inseparable physiological processes: oxygenation and ventilation. SpO2 tells you whether the blood is carrying enough oxygen; end-tidal CO2 (EtCO2) tells you whether the patient is ventilating effectively enough to eliminate carbon dioxide. Understanding what each measurement reflects – and crucially, what it does not – is what separates competent monitoring from safe monitoring.
This guide covers the physiology behind both modalities, how to interpret values and waveforms, where common errors occur, and how to apply both tools across the clinical scenarios you will encounter most often: post-intubation confirmation, procedural sedation, CPR, and ventilated ICU patients. Twenty NCLEX high-yield tips close the article.
Pulse oximetry: how it works and what it measures
The physics of SpO2
A pulse oximeter probe contains two light-emitting diodes that transmit light at 660 nm (red) and 940 nm (infrared) through a vascular tissue bed to a photodetector. Oxyhemoglobin and deoxyhemoglobin absorb these two wavelengths at different ratios. By comparing the pulsatile (arterial) component of absorbance at both wavelengths, the device calculates the percentage of hemoglobin that is bound to oxygen – the functional arterial oxygen saturation (SpO2).
The key distinction nurses must understand is functional vs fractional saturation. Functional saturation is the percentage of hemoglobin capable of carrying oxygen that is saturated. Fractional saturation is the percentage of total hemoglobin that is saturated – and total hemoglobin includes dyshemoglobins like carboxyhemoglobin (COHb) and methemoglobin (MetHb). Standard two-wavelength pulse oximeters measure functional saturation only, which is why carbon monoxide poisoning produces a falsely normal SpO2 (see Limitations below).
The oxyhemoglobin dissociation curve
The relationship between SpO2 and the partial pressure of oxygen (PaO2) is not linear – it follows the S-shaped oxyhemoglobin dissociation curve. Three inflection points are clinically essential:
- PaO2 ~100 mmHg → SpO2 ~98–99%: Normal healthy adult at rest. Increasing PaO2 above 100 mmHg adds very little additional saturation – the curve has plateaued. This is the “flat” upper portion.
- PaO2 ~60 mmHg → SpO2 ~90%: This is the critical inflection point at the “shoulder” of the curve. Above this point, large drops in PaO2 cause only small drops in SpO2. Below it, the curve steepens sharply – small further drops in PaO2 cause disproportionately large drops in saturation. A patient at SpO2 90% is at the cliff edge, not in a stable zone.
- PaO2 ~40 mmHg → SpO2 ~75%: Normal mixed venous oxygen level – the saturation at which tissues return deoxygenated blood to the right heart.
The curve shifts right (reduced hemoglobin affinity for O2, promotes offloading) with acidosis, hyperthermia, hypercapnia, and increased 2,3-DPG. It shifts left (increased affinity, impairs offloading at tissues) with alkalosis, hypothermia, hypocapnia, fetal hemoglobin, and carbon monoxide poisoning. In clinical practice, the right-shift is a compensatory adaptation during hypoxia; the left-shift in CO poisoning is one reason why tissue hypoxia is so severe despite a normal-appearing SpO2.
For a full discussion of the arterial blood gas values that quantify PaO2 and SaO2 directly, see the guide to ABG interpretation.
Probe placement
The finger is the default site – well-perfused, simple to apply, and comfortable. Place the probe so the light source sits over the nail bed and the emitter and detector face each other directly; follow the manufacturer’s site guidance.
Earlobe: Ear probes respond to a change in saturation more quickly than finger probes (Jubran 2015). Useful during rapid clinical deterioration when you need the most current value, and as an alternative site when the fingers are cold or poorly perfused.
Forehead (reflectance probe): Measures SpO2 via reflection rather than transmission. It reads from the supraorbital artery, which is less affected by vasoconstriction than the digits. In one study of apneic anesthetized patients, the forehead probe detected desaturation to 90% at 94 seconds and the finger probe at 100 seconds (Jubran 2015). Useful in patients with cold, poorly perfused extremities.
Nasal alar / nose: Reflectance probe; used when other sites are inaccessible.
Toe / foot: Used in neonates and infants, or when upper extremity sites are unavailable. Note that toe response time is slower than finger and earlobe.
Plethysmographic waveform quality
Most monitoring systems display a plethysmographic (pleth) waveform alongside the numeric SpO2 value. This waveform represents the pulsatile variation in blood volume as detected by the photodetector. A good signal shows a regular, tall, rounded waveform with a clear dicrotic notch – the small notch on the downstroke that represents aortic valve closure. A poor signal is flat, irregular, or noisy with no identifiable dicrotic notch. Never trust a SpO2 numeric value accompanied by a poor pleth waveform – it is the device reporting noise, not physiology.
When the waveform is poor, reposition the probe, try the earlobe, warm the extremity, or dry the probe site before documenting the reading.
Interpreting SpO2 values
Normal ranges and clinical thresholds
| SpO2 range | Interpretation | Nursing action |
|---|---|---|
| 96–98% | Normal daytime saturation in healthy young adults (BTS 2017); values fall slightly with age | Routine monitoring; ensure good waveform quality |
| 94–98% (standard target) | BTS target range for most acutely ill adults who are not at risk of hypercapnic respiratory failure | Titrate oxygen to stay within the range; patients already in range do not need supplemental oxygen for hypoxemia |
| 88–92% (at-risk target) | BTS target for patients at risk of hypercapnic respiratory failure: COPD, morbid obesity, cystic fibrosis, chest wall deformity, neuromuscular disease, or bronchiectasis with fixed airflow obstruction | Titrate oxygen to the prescribed range pending blood gas results; a reading above the range should prompt a reduction in flow per order – see the section below |
| Below the prescribed target | Hypoxemia relative to the patient's target | Assess the patient and the cause; start or increase oxygen per the prescription; a rising oxygen requirement to stay in range is itself a sign of deterioration and needs provider review |
| <90% | Below the shoulder of the dissociation curve – small further falls in PaO2 now cause large falls in saturation | Immediate oxygen titration; provider notification; ABG if not already done; assess cause |
| <85% | Severe hypoxemia – critical | Urgent escalation; rapid response activation if no provider is present; prepare airway equipment |
Population-specific targets: COPD and chronic hypercapnia
Some patients with COPD and other chronic lung, chest wall, or neuromuscular conditions develop a rising PaCO2 and respiratory acidosis when given too much oxygen. The traditional explanation is loss of “hypoxic drive”, but the British Thoracic Society guideline (BTS 2017, section 6.3.1) ranks the mechanisms differently. Worsening ventilation-perfusion mismatch, as oxygen releases hypoxic vasoconstriction in poorly ventilated lung, is the main cause. Reduced hypoxic drive comes second and contributes only a small part, followed by the Haldane effect, absorption atelectasis, gas density, and rebreathing.
The practical rule is the same either way. For patients at risk of hypercapnic respiratory failure, BTS recommends a target SpO2 of 88–92% pending blood gas results, titrating oxygen to stay within that range. In a prehospital trial of patients with COPD exacerbations, high-flow oxygen carried a mortality of 9% versus 2% with oxygen titrated to 88–92% (Austin 2010). An SpO2 of 97–100% in an at-risk patient on supplemental oxygen should prompt the nurse to reduce the flow rate per the prescription, not feel reassured. Never withhold oxygen from a hypoxemic patient because of hypercapnia concerns: titrate to the target range instead, and if a blood gas shows a normal PaCO2, the target can be reset to 94–98%. A full discussion of oxygen delivery devices and titration is in the guide to oxygen therapy nursing.
Limitations and sources of error
Understanding where pulse oximetry fails is as important as knowing normal values. A falsely normal SpO2 in a deteriorating patient is dangerous; a falsely low reading in a stable patient triggers unnecessary interventions. The table below covers the most clinically significant sources of error.
| Factor | Effect on SpO2 reading | Nursing response |
|---|---|---|
| Dark nail polish (black, purple, dark blue) | Adds measurement error – in ventilated ICU patients the average bias was small (black +1.6%) but individual readings scattered widely (Jubran 2015) | Remove polish with an acetone wipe or move the probe to the earlobe; rotating the finger probe 90° does not remove the error |
| Poor peripheral perfusion (shock, hypothermia, vasoconstriction) | Unreliable or unobtainable – inadequate pulsatile flow for detection | Move probe to earlobe or forehead; warm extremity; consider arterial line for continuous SaO2 |
| Motion artifact | Spurious readings – random high or low values | Verify probe is secure; use motion-tolerant probe if available; interpret in context of clinical picture |
| Anemia (severe) | Falsely normal SpO2 despite low oxygen content – hemoglobin that is present is saturated, but there is very little of it | Check hemoglobin/hematocrit; calculate oxygen delivery (DO2); SpO2 alone does not reflect oxygen content in severe anemia |
| Carbon monoxide poisoning | Falsely normal – COHb and OHb absorb red light similarly, so the device counts COHb as OHb; SpO2 overestimates true saturation, and the gap widens as COHb rises (Hampson 1998) | If CO poisoning suspected, do NOT rely on standard pulse oximetry; order co-oximetry (ABG with co-oximetry panel) which measures COHb directly |
| Methemoglobinemia | SpO2 drifts toward about 85% as MetHb rises and then plateaus there (84–86% at MetHb around 35%), overestimating true saturation when the patient is hypoxemic (Barker 1989) | If methemoglobinemia suspected (exposure to oxidizing agents, dapsone, benzocaine), order co-oximetry; prepare for methylene blue administration per order |
| Dark skin pigmentation | Systematic overestimation of SpO2 in patients with darker skin – in one large analysis, occult hypoxemia (SaO2 <88% despite SpO2 92–96%) occurred in 11.7% of paired readings in Black patients vs 3.6% in White patients (Sjoding 2020) | Have a low threshold for checking an arterial blood gas with co-oximetry, even when SpO2 is in the low-normal range; correlate with clinical signs |
| Venous pulsations | Falsely low readings – elevated venous pressure (tricuspid regurgitation, right heart failure) creates venous pulsation that the device may detect | Evaluate probe site; consider alternative location; correlate with clinical assessment |
| Supplemental O2 masking hypoventilation | SpO2 may remain normal while PaCO2 rises dangerously – patient on O2 can hypoventilate significantly before SpO2 falls | Add capnography in any patient on supplemental O2 where ventilation is a concern (post-op, sedation, opioids); do not rely on SpO2 alone to assess ventilation |
| Excessive ambient light | Spurious readings from fluorescent or fiber-optic light entering the probe | Cover probe with opaque material if bright light source cannot be moved |
Continuous vs spot-check monitoring
Continuous monitoring is indicated for any patient at risk of sudden respiratory deterioration: post-operatively, during and after procedural sedation, on opioid infusions or patient-controlled analgesia, receiving supplemental oxygen for acute illness, or with known cardiorespiratory instability.
Spot-check monitoring (periodic measurement, typically q4h with vital signs) is appropriate for stable patients without active respiratory conditions and not on therapies that suppress ventilation.
The monitoring frequency should escalate with clinical acuity, and the decision to upgrade from spot-check to continuous must be individualized to the patient’s risk profile – it cannot be driven solely by the current SpO2 value.
Capnography: waveform analysis and clinical use
What EtCO2 measures
Capnography measures the concentration of carbon dioxide in exhaled gas, expressed as a partial pressure (mmHg) or percentage. End-tidal CO2 (EtCO2) is the CO2 concentration at the end of a complete exhalation – the point at which exhaled gas most closely approximates alveolar CO2 and, by extension, arterial CO2 (PaCO2).
In a patient with normal lung physiology and normal cardiac output, EtCO2 runs approximately 2–5 mmHg below PaCO2. This gradient exists because exhaled gas is diluted by dead-space ventilation (gas from airways that never participates in gas exchange). Normal EtCO2 is 35–45 mmHg. When pulmonary dead space increases – as in pulmonary embolism, severe COPD, or low cardiac output – the EtCO2-PaCO2 gradient widens. In these patients, EtCO2 underestimates PaCO2 more significantly, and the absolute EtCO2 value alone is less useful than the trend.
Two types of capnography are in clinical use:
- Mainstream capnography: Sensor sits directly in the airway circuit. Fast response, no sample lag. Used on intubated patients.
- Sidestream capnography: Gas is aspirated from the circuit or nasal cannula sampling line to an external sensor. Slightly slower response; works on non-intubated patients via nasal cannula or nasal-oral sampling cannula.
Normal capnogram waveform: four phases
A normal capnogram is rectangular with sloped transitions and has four distinct phases. Every nurse monitoring ventilated patients or patients under sedation should be able to identify these on sight.
Phase I (baseline / anatomical dead space): EtCO2 = 0 (or near 0). This is the early part of exhalation – CO2-free gas from the anatomical dead space (trachea, bronchi) exits before alveolar gas reaches the sensor. Normally flat at zero.
Phase II (expiratory upstroke): EtCO2 rises rapidly as alveolar gas displaces dead-space gas at the sensor. Should be steep and linear on a normal capnogram.
Phase III (alveolar plateau): EtCO2 reaches its maximum and forms a near-horizontal plateau as alveolar gas – relatively uniform in CO2 concentration – is continuously exhaled. The plateau should be flat or only slightly upward-sloping. The value at the end of Phase III is the EtCO2 reading. The slight upward slope normally seen reflects heterogeneity in alveolar emptying times.
Inspiratory downstroke (labeled phase 0 in Kodali’s nomenclature; some texts call it phase IV): CO2 rapidly drops back to zero as fresh inspired gas flushes the sensor. Should be steep and return to baseline (zero). The angle between Phase II and Phase III is the alpha angle, which widens in airway obstruction; the near-90° angle between Phase III and the inspiratory downstroke is the beta angle (Kodali 2013).
Abnormal capnogram patterns
| Pattern name | Appearance | Clinical cause | Nursing action |
|---|---|---|---|
| Normal rectangular | Flat Phase I, steep Phase II, flat Phase III plateau, steep inspiratory return to zero | Normal lung physiology | Continue routine monitoring |
| Shark fin waveform | Phase II upstroke is slowed and merges with Phase III – no clear plateau; the whole curve looks like a shark fin or steep slope | Bronchospasm, severe COPD, asthma exacerbation – airways obstruction causes uneven alveolar emptying, so CO2-rich gas arrives at the sensor gradually rather than in a defined wave | Notify provider; administer bronchodilator per order; assess breath sounds; monitor EtCO2 trend for improvement |
| Sloping Phase III plateau | Phase III has a pronounced upward slope rather than being flat | Obstructive disease (COPD, asthma) – heterogeneous alveolar emptying; also seen with incomplete bronchodilator response | Assess for bronchospasm; correlate with auscultation; document for provider |
| Absent / flat waveform (EtCO2 = 0) | No waveform – flat line at zero | Esophageal intubation (most critical); apnea; circuit disconnection; sensor malfunction | Immediately verify ETT placement if intubated – assess for chest rise, auscultate, arrange chest X-ray; if apneic patient, stimulate or initiate rescue breathing; check all circuit connections |
| Sudden drop to near zero | Previously present waveform drops abruptly to near zero mid-monitoring | Airway disconnection, circuit leak, ETT dislodgement, or cardiac arrest (EtCO2 drops with loss of pulmonary perfusion) | Inspect circuit immediately; check ETT position; assess patient responsiveness and pulse; initiate CPR if cardiac arrest confirmed |
| Rising baseline (above zero) | Phase I does not return to zero – baseline drifts upward with each breath | CO2 rebreathing – exhausted CO2 absorber (in anesthesia circuits), insufficient fresh gas flow, or partial circuit obstruction | Check CO2 absorber; increase fresh gas flow; notify anesthesia or respiratory therapy; replace circuit components as needed |
| Sudden rise followed by sustained elevation | EtCO2 rises abruptly and stays up during cardiac arrest with CPR | Return of spontaneous circulation (ROSC) – restored cardiac output dramatically increases CO2 delivery to the lungs | Alert the team leader – check for a pulse at the next rhythm check and prepare post-ROSC care. AHA 2025 treats an abrupt rise (for example, more than 10 mmHg) as a possible sign of ROSC; no absolute cutoff confirms it |
Clinical applications of capnography
Post-intubation confirmation
Continuous waveform capnography is the gold standard for confirming endotracheal tube (ETT) placement in the trachea. When the ETT is correctly placed in the trachea, CO2 is exhaled with each breath and a normal rectangular waveform appears immediately. When the tube is in the esophagus, no CO2 waveform is detected – the EtCO2 reads zero (or falls to near zero within the first few breaths, as any CO2 from carbonated beverages or gastric gas is rapidly washed out).
Do not rely on chest rise, epigastric auscultation, or misting in the tube alone to confirm placement – these are adjuncts, not confirmatory tests. Chest X-ray confirms tip position relative to the carina but cannot be obtained fast enough to serve as the primary confirmation method. The capnogram waveform is immediate, continuous, and unambiguous. Full protocols for ETT placement confirmation are covered in airway management nursing.
Procedural sedation monitoring
Capnography detects hypoventilation and apnea before SpO2 drops, and the lead is longest in a patient breathing supplemental oxygen (Krauss 2007). This detection lag is the most clinically critical reason to use capnography during procedural sedation.
When a sedated patient begins to hypoventilate or becomes apneic, oxygen stored in the lungs keeps the saturation up for a time. In a well-preoxygenated healthy adult, SpO2 can stay above 90% for several minutes of apnea, and the window is much shorter in obese, critically ill, or pediatric patients (Weingart 2012). During this window, capnography has already detected the problem: EtCO2 has stopped cycling (no waveform = apnea) or the value is rising (hypoventilation with CO2 accumulation). Waiting for SpO2 to drop before intervening means the apnea has already gone undetected for some time.
For moderate procedural sedation, the 2018 American Society of Anesthesiologists guideline recommends monitoring ventilation with capnography in addition to clinical observation, unless it is precluded or invalidated by the patient, procedure, or equipment. Assess and document EtCO2 waveform and value at the same intervals as SpO2, heart rate, and blood pressure (ASA 2018 suggests checking responsiveness and blood pressure at 5-minute intervals once moderate sedation is established). See conscious sedation nursing for the full monitoring framework.
CPR and cardiac arrest
During cardiac arrest, cardiac output falls to zero or near zero, and pulmonary blood flow ceases. Without blood flow through the pulmonary vasculature, CO2 cannot be transported to the lungs for exhalation – EtCO2 drops precipitously, typically to less than 10 mmHg or absent waveform.
EtCO2 serves two distinct functions during resuscitation:
-
Compression quality feedback: EtCO2 correlates with cardiac output during CPR. The 2025 AHA adult ALS guideline suggests aiming for an EtCO2 of at least 10 mmHg, and ideally 20 mmHg or higher, as a sign of mechanically adequate compressions. A persistently low value should prompt a check of compression depth, rate, and recoil, although a very low EtCO2 also reflects the underlying arrest and carries a poor prognosis (Sanders 1989).
-
ROSC detection: When spontaneous circulation returns, cardiac output is restored, CO2 is rapidly transported to the lungs, and EtCO2 rises sharply (Falk 1988). AHA 2025 notes that an abrupt rise predicts ROSC but that no absolute cutoff has been identified; a sudden increase of more than about 10 mmHg may indicate ROSC. Confirm with a pulse check at the next planned pause rather than stopping compressions on the EtCO2 change alone.
In intubated adults, an EtCO2 of 10 mmHg or less after 20 minutes of advanced life support may be considered as one part of a multimodal decision to stop resuscitation, never as a stand-alone rule (AHA 2025).
For a full discussion of defibrillation and post-ROSC care, see cardioversion and defibrillation nursing. Capnography protocols in code situations are covered in rapid response and code blue nursing.
Ventilated ICU patients
In mechanically ventilated patients, EtCO2 provides a continuous, non-invasive approximation of PaCO2 and enables real-time ventilator management without requiring serial ABGs for every adjustment. The key clinical uses include:
- Trending PaCO2: Once the EtCO2-PaCO2 gradient is established from an ABG (normal: 2–5 mmHg), EtCO2 trends reliably track PaCO2 changes. A rising EtCO2 in a ventilated patient indicates CO2 retention – increase respiratory rate or tidal volume per order.
- Detecting circuit leaks and disconnections: A sudden drop in EtCO2 to zero or near zero in a ventilated patient is a circuit disconnection or accidental extubation until proven otherwise.
- PEEP adjustment guidance: Changes in EtCO2 following PEEP adjustments can reflect changes in dead-space ventilation and alveolar recruitment.
- Detecting acute changes in pulmonary perfusion: A sudden decrease in EtCO2 in a ventilated patient (without ventilator changes) can indicate a new pulmonary embolism or acute decrease in cardiac output – both reduce CO2 delivery to the lungs.
Ventilated patient management, weaning criteria, and ventilator settings are covered in depth in mechanical ventilation nursing.
Non-intubated patients via nasal sampling cannula
Nasal capnography cannulas deliver supplemental O2 from one port while aspirating exhaled gas samples from nasal prongs for EtCO2 analysis. This enables combined O2 delivery and capnography monitoring in spontaneously breathing patients without an artificial airway. Signal quality is generally good for patients breathing through their nose; mouth-breathing patients require an oral-nasal sampling cannula.
Indications include: procedural sedation, post-operative monitoring in PACU, patients on opioid infusions, any spontaneously breathing patient in whom ventilation adequacy cannot be assumed from SpO2 alone.
Monitoring during patient transfer
Portable capnography is essential for any high-acuity patient transfer – whether inter-unit (floor to ICU) or inter-facility. During transport, SpO2 and EtCO2 should both be monitored continuously, with waveform display confirmed prior to leaving the originating unit. Intubated patients are at particular risk during transport for inadvertent extubation, and a flat EtCO2 waveform on a transport monitor is the earliest detectable sign. Document both values and waveform quality on the transfer handoff record.
When using bag-valve-mask ventilation during transport, capnography confirms each ventilation is reaching the lungs. For BVM technique and circuit management, see bag-valve-mask nursing.
Using SpO2 and EtCO2 together
Pulse oximetry and capnography monitor two independent physiological processes. SpO2 measures oxygenation – whether hemoglobin is carrying oxygen. EtCO2 measures ventilation – whether CO2 is being eliminated. Each has blind spots that the other covers.
What SpO2 detects that capnography misses: Hypoxemia in patients with adequate ventilation – hypoxia from diffusion impairment, V/Q mismatch, or low FiO2 in a patient breathing adequately. Capnography can be completely normal while the patient is severely hypoxemic.
What capnography detects that SpO2 misses: Hypoventilation and apnea in patients on supplemental oxygen, rising CO2 (hypercarbia) before it causes SpO2 change, esophageal intubation, circuit disconnection, ROSC during CPR, inadequate chest compressions. A patient who is apneic for 90 seconds and well pre-oxygenated may still show SpO2 of 97–98%; their capnogram will be flat.
Scenario-based interpretation
Post-operative patient on 3 L/min nasal cannula, SpO2 97%, EtCO2 rising from 40 to 55 mmHg, respiratory rate dropping from 14 to 8 breaths/min: This patient is hypoventilating. SpO2 is reassuringly normal and will remain so for another minute or more – the supplemental O2 is masking the problem. Capnography has caught it. Stimulate the patient, reduce opioid if ordered, notify provider, prepare for reversal agent if respiratory depression is opioid-induced.
Sedated patient undergoing colonoscopy, SpO2 95%, EtCO2 flat waveform for 30 seconds: The patient is apneic. SpO2 has not yet dropped. Capnography has detected the apnea. Stimulate the patient verbally and physically, apply jaw thrust, prepare for BVM ventilation if no response, alert the proceduralist.
Intubated ICU patient, SpO2 88%, EtCO2 48 mmHg: Both oxygenation and ventilation are impaired. SpO2 below target indicates hypoxemia (increase FiO2 or PEEP); elevated EtCO2 indicates CO2 retention (increase respiratory rate or tidal volume per order). This is a combined oxygenation and ventilation problem – both parameters must be addressed. ABG ordered to confirm and guide adjustments.
CPR in progress, SpO2 unobtainable (no perfusion), EtCO2 rises abruptly from 8 to 42 mmHg: Probable ROSC. Check for a pulse and blood pressure at the next pause in compressions. Do not delay pulse check waiting for SpO2 to read – there will not be a SpO2 value until perfusion is restored.
Nursing monitoring protocol
Documentation requirements
Document the following at every monitoring interval:
- SpO2 value and probe location
- Supplemental oxygen in use (device type, flow rate or FiO2)
- EtCO2 value and waveform quality (if capnography in use)
- Respiratory rate
- Patient’s subjective respiratory effort and comfort
- Any alarms triggered and nursing response
For capnography, document whether the waveform is present, morphology (normal rectangular vs abnormal pattern), and any trend changes since the previous documentation point.
Monitoring frequency
Minimum documentation frequencies by context:
- Procedural sedation: Continuous monitoring throughout procedure; document q5 minutes and at baseline, during procedure, and post-procedure recovery intervals per facility policy
- Post-operative (PACU): Continuous monitoring; document per Aldrete/PADSS score intervals (typically q15 min until discharge-ready)
- ICU / mechanically ventilated: Continuous monitoring; document q1h or per unit protocol
- Ward patient on supplemental O2: Minimum q4h; q1–2h if unstable
- Spot-check for stable ward patient: With routine vital signs (typically q4–8h)
Provider notification thresholds
Contact the provider or activate rapid response for:
- SpO2 below the prescribed target range (94–98% for most patients, 88–92% if at risk of hypercapnia) that does not correct with position change and oxygen titration, or a rising oxygen requirement to stay in range
- A new EtCO2 below about 30 mmHg (hyperventilation, possible alkalosis or reduced perfusion) or above about 50 mmHg (significant hypoventilation or CO2 retention); these are common protocol values, so use your facility’s thresholds
- Sudden waveform change – particularly absent waveform in a previously intubated patient
- Any upward trend in EtCO2 with concurrent decrease in respiratory rate (evolving hypoventilation)
- SpO2 drop that is occurring rapidly rather than gradually, regardless of absolute value
Facility protocols may specify tighter thresholds for specific patient populations (post-cardiac surgery, neurological monitoring). Always follow the more conservative threshold – facility protocol or clinical judgment – whichever requires earlier escalation.
Patient and family education
Explaining the monitor
Patients and families frequently become anxious about monitoring equipment, particularly when alarms sound. A brief, clear explanation at the start of monitoring reduces anxiety and improves cooperation.
For the pulse oximeter: “This small clip on your finger measures the oxygen level in your blood. It shines a light through your skin – it doesn’t hurt and doesn’t draw blood. We’ll see a number on the screen showing your oxygen percentage. If the number drops too low or the signal is lost, an alarm sounds so we can check on you.”
For capnography via nasal cannula: “This is a special nasal cannula that both delivers oxygen and samples the air you breathe out. It measures the carbon dioxide you exhale, which tells us how well you’re breathing. It doesn’t go into your airway – it just sits at your nose.”
Managing alarm anxiety
Reassure patients that alarms are precautionary – most are triggered by movement, probe position, or a brief interruption and do not indicate a clinical problem. However, explicitly tell the patient: “If the alarm sounds and it bothers you, press your call button – do not remove the probe yourself, because we need to see the signal to know everything is fine.”
For families: explain that a monitoring alarm does not automatically mean something is wrong. Staff will assess and determine whether it represents a true clinical change or a technical artifact. Encourage them to notify the nurse rather than attempt to adjust or silence equipment themselves.
NCLEX review: 20 high-yield tips
| Tip # | Clinical pearl |
|---|---|
| 1 | Know the target ranges: 94–98% for most acutely ill adults and 88–92% for patients at risk of hypercapnic respiratory failure (BTS 2017). Below 90%, the oxyhemoglobin dissociation curve is on its steep portion and small further drops in PaO2 cause large drops in saturation. |
| 2 | Normal SpO2 in a healthy young adult is about 96–98%. A PaO2 of about 60 mmHg corresponds to an SpO2 of about 90% – the shoulder of the oxyhemoglobin dissociation curve. |
| 3 | COPD target SpO2 is 88–92% pending blood gases. Excess oxygen raises PaCO2 mainly by worsening V/Q mismatch; loss of hypoxic drive plays a smaller part (BTS 2017). Titrate oxygen to the range – never withhold it from a hypoxemic patient. |
| 4 | Dark nail polish – especially black, purple, or dark blue – makes SpO2 readings less reliable. Remove polish with acetone or move the probe to the earlobe. |
| 5 | Carbon monoxide poisoning: standard pulse oximetry reads carboxyhemoglobin (COHb) as oxyhemoglobin – SpO2 appears falsely normal (often 98–100%). Order co-oximetry (ABG with co-oximetry) to measure COHb directly. |
| 6 | Methemoglobinemia: SpO2 drifts toward about 85% and plateaus there as MetHb rises, whatever the true saturation. Suspect with exposure to dapsone, benzocaine, or other oxidizing agents. Confirm with co-oximetry. |
| 7 | SpO2 does not detect hypercarbia (elevated CO2). A patient can have SpO2 of 98% and a PaCO2 of 80 mmHg simultaneously – SpO2 measures oxygenation, not ventilation. |
| 8 | Supplemental oxygen masks hypoventilation. After preoxygenation, SpO2 can stay normal for several minutes of apnea. Use capnography to detect apnea before SpO2 falls. |
| 9 | Normal EtCO2 range is 35–45 mmHg. Values outside this range indicate hypoventilation (>45 mmHg, CO2 retention) or hyperventilation (<35 mmHg, excessive CO2 washout). |
| 10 | Capnography is the gold standard for confirming ETT tracheal placement. A normal rectangular waveform confirms tracheal placement; a flat waveform (EtCO2 = 0) after attempted intubation indicates esophageal intubation until proven otherwise. |
| 11 | Capnography detects apnea before SpO2 drops, especially in a patient receiving supplemental oxygen – making it the earlier warning tool during procedural sedation. |
| 12 | EtCO2 during CPR: aim for at least 10 mmHg, ideally 20 mmHg or more (AHA 2025). A persistently low value should prompt a check of compression depth and rate. |
| 13 | ROSC during CPR: an abrupt, sustained rise in EtCO2 (for example, more than 10 mmHg) may signal ROSC. AHA 2025 identifies no absolute cutoff, so confirm with a pulse check at the next planned pause. |
| 14 | Shark fin capnogram = bronchospasm. The slow, sloping waveform with no defined plateau is caused by uneven airway obstruction that delays alveolar gas reaching the sensor. Administer bronchodilator per order. |
| 15 | EtCO2 is typically 2–5 mmHg lower than PaCO2 in patients with normal lungs and cardiac output. The gradient widens with increased dead space (pulmonary embolism, low cardiac output, severe COPD). |
| 16 | A rising baseline on the capnogram (Phase I does not return to zero) indicates CO2 rebreathing – check CO2 absorber in anesthesia circuits or increase fresh gas flow. |
| 17 | Earlobe and forehead probes respond to saturation changes faster than finger probes. Use them in low-perfusion states or when you need the most current oxygenation status. |
| 18 | Severe anemia can produce a falsely reassuring SpO2. If hemoglobin is very low, SpO2 may read 98% because the small amount of hemoglobin present is fully saturated – but total oxygen content (CaO2) is critically reduced. Check Hgb/Hct alongside SpO2. |
| 19 | Poor pleth waveform = unreliable SpO2. If the waveform on the monitor is flat or irregular, the numeric value cannot be trusted. Reposition the probe, warm the extremity, or move to an alternative site before documenting or acting on the reading. |
| 20 | SpO2 and EtCO2 monitor different physiological processes and each has blind spots the other covers. SpO2 detects oxygenation failure; EtCO2 detects ventilation failure. In any high-acuity patient, both together provide a more complete respiratory picture than either alone. |
Summary
Pulse oximetry and capnography are complementary monitoring modalities that provide real-time assessment of two separate and essential respiratory functions. SpO2 is the nurse’s window into oxygenation – whether hemoglobin is carrying sufficient oxygen to meet metabolic demand. EtCO2 is the window into ventilation – whether CO2 is being eliminated effectively and the airway is patent.
Competent use of these tools requires more than reading the numeric values. It requires understanding where each modality can mislead – from nail polish and poor perfusion affecting SpO2, to the EtCO2-PaCO2 gradient widening in disease states – and recognizing the waveform patterns that carry the most clinical urgency: the absent waveform of esophageal intubation, the shark fin of bronchospasm, the rising EtCO2 that heralds ROSC. Used together and interpreted in context, they form one of the most powerful surveillance systems available to the bedside nurse.
References
- Jubran A. Pulse oximetry. Crit Care. 2015;19(1):272. doi:10.1186/s13054-015-0984-8
- Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383(25):2477–2478. doi:10.1056/NEJMc2029240
- Amalakanti S, Pentakota MR. Pulse oximetry overestimates oxygen saturation in COPD. Respir Care. 2016;61(4):423–427. doi:10.4187/respcare.04435
- Practice guidelines for moderate procedural sedation and analgesia 2018: a report by the American Society of Anesthesiologists Task Force on Moderate Procedural Sedation and Analgesia, et al. Anesthesiology. 2018;128(3):437–479. doi:10.1097/ALN.0000000000002043
- Kodali BS. Capnography outside the operating rooms. Anesthesiology. 2013;118(1):192–201. doi:10.1097/ALN.0b013e318278c8b6
- Bhende MS, Thompson AE. Evaluation of an end-tidal CO2 detector during pediatric cardiopulmonary resuscitation. Pediatrics. 1995;95(3):395–399. doi:10.1542/peds.95.3.395
- Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: Adult Advanced Life Support: 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 Suppl 2):S538–S577. doi:10.1161/CIR.0000000000001376
- Silvestri S, Ralls GA, Krauss B, et al. The effectiveness of out-of-hospital use of continuous end-tidal carbon dioxide monitoring on the rate of unrecognized misplaced intubation within a regional emergency medical services system. Ann Emerg Med. 2005;45(5):497–503. doi:10.1016/j.annemergmed.2004.09.014
- Krauss B, Hess DR. Capnography for procedural sedation and analgesia in the emergency department. Ann Emerg Med. 2007;50(2):172–181. doi:10.1016/j.annemergmed.2006.10.016
- O’Driscoll BR, Howard LS, Earis J, Mak V; British Thoracic Society Emergency Oxygen Guideline Group. BTS guideline for oxygen use in adults in healthcare and emergency settings. Thorax. 2017;72(Suppl 1):ii1–ii90. doi:10.1136/thoraxjnl-2016-209729
- Austin MA, Wills KE, Blizzard L, Walters EH, Wood-Baker R. Effect of high flow oxygen on mortality in chronic obstructive pulmonary disease patients in prehospital setting: randomised controlled trial. BMJ. 2010;341:c5462. doi:10.1136/bmj.c5462
- Falk JL, Rackow EC, Weil MH. End-tidal carbon dioxide concentration during cardiopulmonary resuscitation. N Engl J Med. 1988;318(10):607–611. doi:10.1056/NEJM198803103181005
- Sanders AB, Kern KB, Otto CW, Milander MM, Ewy GA. End-tidal carbon dioxide monitoring during cardiopulmonary resuscitation: a prognostic indicator for survival. JAMA. 1989;262(10):1347–1351. doi:10.1001/jama.1989.03430100081033
- Eberhard P. The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg. 2007;105(6 Suppl):S48–S52. doi:10.1213/01.ane.0000278642.16117.f8
- Barker SJ, Tremper KK, Hyatt J. Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry. Anesthesiology. 1989;70(1):112–117. doi:10.1097/00000542-198901000-00021
- Hampson NB. Pulse oximetry in severe carbon monoxide poisoning. Chest. 1998;114(4):1036–1041. doi:10.1378/chest.114.4.1036
- Weingart SD, Levitan RM. Preoxygenation and prevention of desaturation during emergency airway management. Ann Emerg Med. 2012;59(3):165–175.e1. doi:10.1016/j.annemergmed.2011.10.002